Rohde & Schwarz NGL202 Power Supply

Due to their high accuracy and fast load recovery time, the R&S®NGL200 power supply series is perfect for challenging applications. Their two-quadrant architecture allows them to function both as a source and a sink to simulate batteries and loads. Their short recovery times enable them to handle fast load changes that occur for example when mobile communications devices switch from sleep mode to transmit mode.


Features & Benefits


Optimized Low Recovery Time


Due to the optimized load recovery time of < 30 µs with minimal overshoot during challenging load conditions, the R&S®NGL200 instruments are perfect when testing IoT and other battery-powered devices which require very little current in sleep mode and abruptly increase current when switching to transmit mode.

6 1/2 Digit Resolution 


With up to 6½ digit resolution when measuring voltage, current and power, the R&S®NGL200 series is optimal for characterization of devices with low standby power consumption and high current in full load operation. It can even replace an additional DMM in many applications.



Minimal Ripple and Noise


To supply interference-free voltage to sensitive designs such as complex semiconductors and to support the development of power amplifiers and MMICs.



Two-Quadrant Operation


The two-quadrant architecture of the R&S®NGL200 power supplies allows them to function both as a source and a sink and simulate batteries or loads. The power supply automatically switches from source to sink mode. As soon as the externally applied voltage exceeds the set nominal voltage, current flows into the power supply. This is indicated by a negative current reading.


Save up to 25% when you trade in your qualifying instrument


NGL202 was $4,075. now $3,625


Add Items to Cart



Specifications
All data is valid at +23°C (-3°C/+7°C) after 30 minutes warm-up time.

Outputs The channel outputs are galvanically isolated and not connected to the ground  
Number of output channels NGL201 1
  NGL202 2
Total output power NGL201

max. 60 W

  NGL202 max. 120 W
Maximum output power per channel   60 W
Output voltage per channel  

0 V to 20 V

Maximum output current per channel ≤ 6 V output voltage 6 A
  > 6 V output voltage 3 A
Maximum voltage in serial operation NGL202 40V
Maximum current in parallel operation NGL202, ≤ 6 V ouput voltage 12 A
  NGL202, > 6 V output voltage 6 A
Adjustable output impedance   -50 mΩ to 100 Ω
Increments   1 mΩ
Voltage ripple and noise 20 Hz to 20 MHz < 500 µV (RMS)
Current ripple and noise 20 Hz to 20 MHz < 2 mV (peak-to-peak) (meas.)
Electronic Load    Yes, NGL202: both channels
Maximum sink power NGL201 60 W
  NGL202 120 W (60 W per channel) *
Maximum sink current per channel   3 A
Sink modes   Constant voltage, constant current, contant resistant
Load Regulation  Load change: 10% to 90%  
Voltage ± (% of the output + offset) < 0.01% + 0.5 mV
Current ± (% of the output + offset) < 0.01% + 0.1 mA
Load recovery time regulation to within ±20 mV of the set voltage < 30 µs (meas.)
Programming resolution    
Voltage ± (% of the output + offset) < 0.01% + 3 mV
Current ± (% of the output + offset) < 0.01% + 2 mA


* Time limited at an operating temperature of > 30°C and total power > 90 W.